Physical Property Measurement System (PPMS)
Instrument Administrator: Yu-Ting Lin
Location: RCCI 3F R315
Contact information: yutinglin@as.edu.tw
Introduction
The Physical Property Measurement System (PPMS) is equipped with a liquid helium-free closed-loop system, utilizing a two-stage pulse tube cryocooler to provide a low-vibration cryogenic environment for both superconducting magnets and sample measurements. This system ensures continuous low-temperature control and supports precise magnetic field and temperature sweeping modes. The PPMS features an accurate temperature control system and external magnetic field application, making it a versatile platform for various measurements. The system offers a magnetic field range of 0–14 Tesla and a temperature range of 1.8 K to 300 K.
The main unit generates a high vacuum (<1E-4 torr) through a cold pump, allowing the PPMS DynaCool to support future upgrades for various measurement options, including magnetic, thermal, and electrical properties such as magnetoresistance, differential resistance, Hall coefficient, IV characteristics, critical current, AC susceptibility, magnetic hysteresis, thermal magnetic curves, specific heat, thermoelectric effects, See beck coefficient, thermal conductivity, expansion coefficient, and morphology characterization.
Currently, the center only has the DC Resistivity Measurement Module, which enables DC resistivity measurements by varying the temperature under a fixed magnetic field or by varying the magnetic field under a fixed temperature.
Brand/Model: Quantum Design / DynaCool
DC Resistivity:
Resistance Range: 10 μΩ to 5 MΩ
Sensitivity: 15 nV RMS typical
Current Range: 10 nA to 8 mA
Compliance Voltage: 4 V, maximum
Frequency: 5 Hz square wave
Temperature Range: 1.8 K〜300 K
Magnetic Field Range: 0〜14 T
Service
1. Sample Sizes: Samples must be smaller than 1.2 × 1.2 cm.
2. Measurement Options:
2.1. Simultaneous measurement of up to 3 channels.
2.2. DC resistivity measurements with temperature variation under a fixed magnetic field.
2.3. DC resistivity measurements with magnetic field variation under a fixed temperature.
3. Wire Bonder Service: Wiring samples to the sample puck is available.
4. Long-Term Measurements: Users should prepare their sample puck for frequent use.
*Sample Puck Information:
Product Name:DC Resistivity Sample Puck
Model: 8084-109