Cryogenic magneto-optical system
Instrument Administrator: Shin-Wen Chen
Location: RCCI 3F R313
Contact information: qws5718340@gate.sinica.edu.tw
Introduction
The attoDRY2100 is a premier automated research platform designed specifically for quantum optics and condensed matter physics. The system combines a 1.8 K cryogenic environment, a ±9 T superconducting magnetic field, and high-resolution scanning optical technology. Utilizing its patented vibration-isolation structure and dedicated low-temperature high-NA objectives, it delivers high-stability measurements under extreme physical environments.
Additionally, the integrated Single Quantum near-infrared superconducting nanowire single-photon detector (IR SNSPD) provides highly sensitive detection for faint quantum signals in telecommunication bands. It is a vital tool for studying near-infrared single-photon dynamics and quantum light source characteristics.
Specifications
- Temperature Range: 1.8 K~300 K
- Magnetic Field Range: X-axis< 3T, Z-axis< 9T
- Cryogenic High-NA Objective Wavelengths: 700 nm, 900 nm, and 1300 nm
- Applicable Spectral Range: 600 nm~1700 nm
- Sample Stage Travel Range: 5mm x 5mm
- Light Sources:
|
Model |
picoEmerald S |
Chameleon Ultra II |
SuperK Fianium |
Lumentum UN162 |
|
Laser Type |
Picosecond OPO |
Femtosecond Ti:Sapphire |
Supercontinuum (White Light) |
Industrial Picosecond Solid-State |
|
Pulse Width |
~2 ps |
< 140 fs |
~5 - 50 ps |
~10 ps |
|
Wavelength Range |
700 - 1950 nm |
680 - 1080 nm |
390 - 2400 nm |
1064 nm |
Service
- Cryogenic Magneto-Optical Spectroscopy (Variable temperature / Variable magnetic field)
- Near-Infrared Single-Photon Dynamics Measurements (Across 800 nm~ 1600 nm NIR / Telecommunication bands)
Examples: Time-Resolved Photoluminescence (TRPL), Second-Order Correlation Function g(2), and other single-photon experimental measurements.
Sample Specifications:
1. Size: ≤ 10 mm × 10 mm
2. Thickness: ≤ 5 mm
3. All samples must be properly sealed and packaged by the user